ViSolid®: Optical sensors for the in-situ use to measure suspended solids via scattered light and direct back-scattering with ultrasonic cleaning system:
- Ultrasonic cleaning without mechanics
- Extremely low-maintenance
- Highly accurate factory calibration
- High operational safety (sensor check function)
With higher particle concentrations, the procedures of turbidity measurement are only suitable in some cases; here, you will need special measuring techniques to determine the concentration. WTW uses two procedures, which are used depending on the suspended solids concentration: With low concentrations, a scattered light measurement is used; for higher concentrations, the direct back-scattering yields optimal results.
The sea water model includes the sensor for use in special media: All wetted parts are made of titanium and plastic and are therefore extremely resistant to corrosion.
ViSolid®: Optical sensors for the in-situ use to measure suspended solids via scattered light and direct back-scattering with ultrasonic cleaning system:
- Ultrasonic cleaning without mechanics
- Extremely low-maintenance
- Highly accurate factory calibration
- High operational safety (sensor check function)
With higher particle concentrations, the procedures of turbidity measurement are only suitable in some cases; here, you will need special measuring techniques to determine the concentration. WTW uses two procedures, which are used depending on the suspended solids concentration: With low concentrations, a scattered light measurement is used; for higher concentrations, the direct back-scattering yields optimal results.
The sea water model includes the sensor for use in special media: All wetted parts are made of titanium and plastic and are therefore extremely resistant to corrosion.
Model | ViSolid® 700 IQ | ViSolid® 700 SW* |
Order No. | 600012 | 600013 |
* SW: Sensor as sea water model (with plastic arming (POM) | ||
Measuring Method | Procedure for measuring scattered light | |
Measuring Range | g/1 Si0z: 0 ... 300 g/1 Si0z | |
Resolution | g/1 Si0z: Automatic according to measuring range 0.1 mg/1 ... 1 g/1 | |
Calibration | Typical sludge characteristic stored: matrix type 1, matrix type 2 Calibration by user: adjustment via correction factor, 1 - point or | |
Cleaning System | Ultrasound cleaning system | |
SensCheck | Contamination detection of optical window; failure of cleaning system | |
Pressure Resistance | 10 bar (incl. sensor connection cable) | |
Ambient Conditions | Operating temperature: 32 ... 140°F (0 ... 60°C): (Overheating protection); | |
Electrical Connections | 2-wire shield cable with quick fastener to sensor | |
Electromagnetic | EN 61326, Class B, FCC Class A; Intended for indispensable operation | |
Compatibility | ||
Certifications | CE | |
Mechanical | Measuring Window: Sapphire; | Measuring Window: Sapphire; |
Weight (without cable) | Approx. 2.18lb (900 g) | Approx. 3.13lb (1420 g) |
Warranty | 2 years for defects in quality | |